Published April 5, 2022 | Version v1
Journal article Open

Providing Enhanced Migration Time Reproducibility with a High-Voltage-Compatible Flow Sensor for Capillary Electrophoresis

Description

In capillary electrophoresis (CE), analyte identification is primarily based on migration time, which is a function of the analyte's electrophoretic mobility and the electro-osmotic flow (EOF). The migration time can be impacted by the presence of parasitic flow from changes in temperature or pressure during the run. Presented here is a high-voltage-compatible flow sensor capable of monitoring the volumetric flow inside the capillary during a separation with nL/min resolution. The direct measurement of both flow and time allows for compensation of flow instabilities. By expressing the electropherogram in terms of signal versus electromigration velocity instead of time, it is possible to improve the run-to-run reproducibility up to 25×.

Notes

+ This version of the article has been accepted for publication, after peer review (when applicable) and is subject to ACS Publications AM terms of use, but is not the Version of Record and does not reflect post-acceptance improvements, or any corrections. The Version of Record is available online at: https://pubs.acs.org/doi/10.1021/acs.analchem.2c00038 + ID der Publikation: hslu_89481 + Art des Beitrages: Wissenschaftliche Medien + Jahrgang: 94 + Sprache: Englisch + Letzte Aktualisierung: 2023-03-29 14:38:28

Files

Kehl_et_al._2022_-_Providing_Enhanced_Migration_Time_Reproducibility_with_a_High-Voltage-Compatible_Flow_Sensor_for_Capillary_Electrophoresis_no_logo.pdf

Additional details

Related works

Is identical to
10.1021/acs.analchem.2c00038 (DOI)