A.3.3.1 -Characterization of flow rate, volume, type of liquid, bolus and other related characteristics of an insulin pump, a syringe pump, a pain pump and a Infusion device analiser

  • 1. Niemann
  • 2. Kjeldsen
  • 3. Batista
  • 4. Gersl
  • 5. Bissig
  • 6. Mills
  • 7. Morgan
  • 8. Ogheard
  • 9. Büker
  • 10. Kartmann
  • 11. Lötters

Description

The aim of this report is to present all the test results obtained in task A3.1.2  of project MeDDII, mainly characterise the flow rate, delivered volume, type of liquid, pressure, bolus size and other relevant characteristics of an insulin pump, a syringe pump, a pain pump and an Infusion device analiser (IDA). A protocol for the test to be performed was developed prior to the beginning of the task. Several laboratories participated in this test. All the results are presented in tables but only a part of them are represented in graphics.