Repeatability and Reproducibility of Pterygium Grading Based on Clinical Translucence Appearance
Creators
- 1. Department of Optometry and Visual Sciences, Kulliyyah of Allied Health Sciences, Inter-national Islamic University Malaysia (IIUM), Kuantan, Pahang, Malaysia
Description
Background: The goal of this study was to evaluate the repeatability and reproducibility of pterygium grading based on clinical translucence appearances and its reliability in clinical setting.
Methods: A total of 93 primary pterygium eyes from 93 participants who visited a tertiary ophthalmic clinic were recruited in this study. Two (2) ophthalmologists and two (2) optometrists evaluated and graded the 93 primary pterygium images in randomized fashion. Graders were instructed to utilise the clinical translucence appearance of pterygium to grade them into type I, II and III. Repeatability testing was done by a single expert by comparing grading of each image on two separate sessions, with one month interval between sessions. Reproducibility was tested by comparing the grading obtained by both experts and optometrists.
Results: Paired and independent T-test results showed no significance difference between graders for both experts and optometrists’ group (all P > 0.05). Intra-grader and inter-grader reliability estimates were high with intraclass correlation ranging from 0.97 to 0.98 for both groups.
Conclusions: Pterygium clinical grading based on its translucence appearance is reliable and repeatable in clinical setting, easily to be graded, interpreted, and recommended for clinicians with different levels of experience.
Files
JORVC2100102.pdf
Files
(693.2 kB)
Name | Size | Download all |
---|---|---|
md5:105a62576e3e6e0727fe430bf0c7fff8
|
693.2 kB | Preview Download |
Additional details
Related works
- Is cited by
- Peer review: 10.54289/JORVC2100102 (DOI)