Published November 9, 2019 | Version v1
Journal article Open

A novel method for producing NH4+ reagent ions in the hollow cathode glow discharge ion source of PTR-MS instruments

  • 1. IONICON Analytik GmbH, Innsbruck, Austria
  • 2. IONICON Analytik GmbH, Innsbruck, Austria; Institut für Ionenphysik und Angewandte Physik, Universität Innsbruck, Innsbruck, Austria
  • 3. Institut für Ionenphysik und Angewandte Physik, Universität Innsbruck, Innsbruck, Austria; Department of Chemistry, University of Oslo, Oslo, Norway

Description

Proton-transfer-reaction mass spectrometry (PTR-MS) is seeing an increased use of NH4+ ions for the detection of amines and labile oxygenated organic compounds. NH4+ ions are usually generated from ammonia or ammonium-containing chemicals. We herein present a simple method for generating NH4+ reagent ions in the hollow cathode glow discharge ion source that nowadays most proton-transfer-reaction mass spectrometry (PTR-MS) instruments are equipped with. We show that NH4+ ions can be generated in high purity and yield by simply introducing a mixture of water vapor and nitrogen in the ion source. We also show that rapid switching (∼10 s loss of data) between the H3O+ and NH4+ operation modes is possible. Our new method can be easily implemented in most PTR-MS instruments that are currently in use, thus opening the possibility to easily operate the PTR-MS analyzer in the NH4+ chemical ionization mode.

Files

NH4 Paper.pdf

Files (632.0 kB)

Name Size Download all
md5:5e840f533224418f55cfb227d56006a6
632.0 kB Preview Download

Additional details

Funding

IMPACT – Ion-Molecule Processes for Analytical Chemical Technologies 674911
European Commission