There is a newer version of the record available.

Published May 14, 2025 | Version v3
Dataset Open

Supplementary material for "An acoustic lens for displacement-free sectorial inspection of pipes with ultrasonic phased arrays"

  • 1. ROR icon Petrobras (Brazil)
  • 2. ROR icon Universidade Tecnológica Federal do Paraná
  • 3. ROR icon Universidade Federal do Rio de Janeiro

Description

This repository contains the data and source code used to produce the results presented in:

Pires, Gustavo P. and Kalid, Thiago E. and Prado, Tatiana de A. and Costa, Vinicius L. and Pereira, Gabriela R. and Passarin, Thiago A. R. and Pipa, Daniel, An Acoustic Lens for Displacement-Free Sectorial Inspection of Pipes with Ultrasonic Phased Arrays. Available at SSRN: https://ssrn.com/abstract=5224648 or http://dx.doi.org/10.2139/ssrn.5224648

 

 

Abstract

Pipes that transport acidic fluids are prone to pitting corrosion, which causes well-localized thickness loss. The detection and measurement of these small flaws via ultrasonic non-destructive testing (UT) requires a fine sampling of the surface of the pipe, which is time-consuming and expensive, especially in places where the inspections require dedicated infrastructure, such as subsea facilities. In this work, we present an acoustic lens that, used with a UT phased array system capable of beam steering, allows for the inspection of large sections (e.g., 90 degrees) of a submerged pipe without any mechanical displacement. The mathematical formulation is derived from geometrical optics, and algorithmic procedures to compute customized delay laws are provided. Practical aspects of the manufacturing and assembly of the device are also described. Spatial resolutions of a prototypical manufactured lens on three axes are assessed on a 140 mm-diameter pipe and compared to those obtained with a moving single-element transducer, showing competitive resolutions. Measurements of the remaining thicknesses of 1.5 mm-diameter machined flaws performed with the device provided a mean absolute error of 19 % of one wavelength and maximum absolute error of 51 % (λ=1.18 mm). The results demonstrate the potential of the proposed concept to replace single-element UT,  significantly reducing inspection time by removing displacement otherwise necessary during an axial sweep.

License

All Python source code (.py) is made available under the MIT license. You can freely use and modify the code, without warranty, so long as you provide attribution to the authors. See LICENSE-MIT.txt for the full license text.

Figures and data produced as part of this research are available under the Creative Commons Attribution 4.0 License (CC-BY). See LICENSE-CC-BY.txt for the full license text.

Files

code.zip

Files (19.4 GB)

Name Size Download all
md5:51df5d48bb3449550c88ba0e81b5d1ec
3.6 MB Preview Download
md5:baa88dfb21b5bfa14b2fd71271cde9c7
19.4 GB Preview Download

Additional details

Related works

Funding

Petrobras (Brazil)
AUSPEX2 2022/00436-2
National Agency of Petroleum, Natural Gas and Biofuels
PRH 21
National Council for Scientific and Technological Development
316234/2021-4

Software

Repository URL
https://github.com/thiagokalid/acoustic-lens-pipe-inspection
Programming language
Python
Development Status
Active