The basic scheme of the multi-target evaluation (MTE) framework
Description
The following supporting information can be downloaded at: https://www.mdpi.com/article/... /supplementary_MTE.mp4. Video S1: MTE. The basic scheme of the multi-target evaluation (MTE) framework is shown in this video for a specific TR3 size, using the HPA for mapping SAR10g,max. For MTE, hypothetical TRs (borderlines marked in black and green) are moved across the entire brain slice for the evaluation of the quality of targeted RF heating instead of using the conventional approach of defining a limited number of specific but arbitrary TRs. EMF simulations were performed for each TR position across a 26x27 rectangular grid with 702 points (step size = 7mm). For evaluation, the metric success score was used, which represents the percentage of the number of locations with SAR10g,max>40 w/kg over the total number of locations. Video S2-S9: APA/HPA_TR1-TR4_MTE
Files
supplementary_MTE.zip
Files
(152.4 MB)
Name | Size | Download all |
---|---|---|
md5:6a52d4a50f93e30e06fd03fe83f8d30c
|
152.4 MB | Preview Download |