Published April 13, 2023 | Version v1

A parallel-beam wavelength-dispersive X-ray emission spectrometer for high energy resolution in-air micro-PIXE analysis

  • 1. Jožef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia
  • 2. Biotechnical Faculty, University of Ljubljana, Jamnikarjeva 101, 1000 Ljubljana, Slovenia

Description

A new parallel-beam wavelength dispersive (PB-WDS) X-ray emission spectrometer was constructed at the external proton beamline at the Microanalytical Centre of the Joˇzef Stefan Institute in Ljubljana. The spectrometer combines polycapillary X-ray optics for efficient X-ray collection with diffraction on a flat crystal analyzer and achieves energy resolution in the eV range. The whole set-up is enclosed within a He bag to be able to operate in the tender X-ray energy range. The basic design is described together with the results of characterization measurements yielding the main operation characteristics. Finally, an application for the micro-PIXE analysis of biological tissue is demonstrated exploiting both spatial and energy resolution of the new set-up.

Files

d3ja00019b.pdf

Files (1.0 MB)

Name Size
md5:3538e9628bfd9f5bf662b1df4ab60e7c
1.0 MB Preview Download

Additional details

Identifiers

Funding

European Commission
RADIATE - Research And Development with Ion Beams – Advancing Technology in Europe 824096