Published February 24, 2022 | Version v1
Dataset Open

Electromagnetic immune phosphor-tipped fibre-optic thermometers

  • 1. National Physical Laboratory
  • 2. Danish Technological Institute
  • 3. Aalborg University Hospital

Description

Calibration traceability can be broken where there are significant, unquantified uncertainties. Thermometry in harsh environments using electrical sensors, such as thermocouples or resistance thermometers, can be unpredictably affected by electro-magnetic sources. For this application, phosphor based temperature sensors with fibre-optic connection have been made following two different approaches – phosphor decay time changes and phosphor emission spectral changes – and tested by bombardment with high energy electrons and by measurements in large magnetic fields. Both sets of tests showed good immunity to exposure, and with magnetic field tests significantly better than a thermocouple. Phosphor thermometry therefore has potential to retain traceability in situations where conventional sensors might fail.

Notes

https://iopscience.iop.org/article/10.1088/1361-6501/ac5440

Files

Files (865.8 kB)

Name Size Download all
md5:df1fff1312cc00e8387673bc7f579ae6
279.6 kB Download
md5:efc30316c4c8e6869e757df8798dc4bc
19.6 kB Download
md5:bd18ab47549459051500d208a1c91789
18.4 kB Download
md5:b3d684c78f1fd2ca8cb708a5822be634
4.9 kB Download
md5:b0f04bc55903f393b410678211a42458
528.1 kB Download
md5:a5029203fa608f2ad304840d6273555d
13.3 kB Download
md5:20ce615d952a66ceff0f72065aa4fa0e
1.9 kB Download