Multi-Line Gamma-Ray Spectrometer Performance of a Si(Li) Detector Stack
Description
Experimental data is presented which for the first time displays multi-line spectrometer performance of a Si(Li) detector stack at elevated temperature. The stack consists of four elements, each with a 2 cm diameter active area. Ba-133 and Ag-110m spectra are obtained using various techniques to enhance the peak-to-background ratio. Spectral data are shown as a function of temperature (94 K less than or = T less than or = 230 K) using optimized peak shaping.