Polarographic investigation of the catalytical adsorptive complex behavior of trace iron

Description

Department of Chemistry, Shandong University, Jinan, P.R. China

Manuscript received 23 December 1998, accepted 28 December 1998

In HAc-NaAc (1 : 1) buffer solution, FeIII-5-Br-PADAP-NO3 system sensitively presents a reductive wave at a banging mercury drop electrode (HMDE). The reductive wave is a catalytical adsorptive complex wave. In 0.05 mol dm-3 HAc-NaAc and 4.0 x 10-6 mol dm-3 5-Br-PADAP solution, the second derivative peak height of the reductive wave is linearly proportional to the concentration of FeIII in the range from 1.0 x 10-10 to 1.0 x 10-7 mol dm-3. The detection limit is 6.0 x 10-11 mol dm-3. The method has been applied to determine trace amount of iron samples of pig meat and tap water.