Published September 29, 2021
| Version v1
Other
Open
Trace metal analysis in colloidal silica by ICP-MS after sample dilution and addition of HF
Description
The method is used for the analysis of elements in colloidal silica with ICP-MS (Agilent 7700). The sample is diluted by ultrapure water to a concentration of 0,2-0,3% SiO2, and small volumes of hydrofluoric acid (HF) and nitric acid (HNO3) are added to dissolve the silica. The sample is thereafter directly analyzed by ICP-MS.
Files
8_SOP_Nouryon_ICP-MS .pdf
Files
(356.4 kB)
Name | Size | Download all |
---|---|---|
md5:b1dc2bf3fe506842e20921238edf1df4
|
356.4 kB | Preview Download |