Published January 5, 2016
| Version v1
Poster
Open
Verification of Absolute Calibration of Quantum Efficiency for LSST CCDs
Authors/Creators
- 1. Wayne State University / Brookhaven National Laboratory
- 2. SLAC National Laboratory
- 3. Wayne State University
- 4. Purdue University
- 5. Brookhaven National Laboratory
- 6. SLAC National Accelerator Laboratory
- 7. Harvard University
Description
We describe a system to measure the Quantum Efficiency in the wavelength range of 300nm to 1100nm of 40x40 mm n-channel CCD sensors for the construction of the 3.2 gigapixel LSST focal plane. The technique uses a series of instruments to create a very uniform flux of photons of controllable intensity in the wavelength range of interest across the face of the sensor. This allows the absolute Quantum Efficiency to be measured with an accuracy in the 1% range. This system will be part of a production facility at Brookhaven National Lab for the basic components of the LSST camera.