  • 1. 1Institute of Neuroscience and Medicine (INM-1), Forschungszentrum Jülich, Jülich 52425, Germany 2JARA-Institute for Quantum Information, RWTH Aachen University, Aachen 52056, Germany 3Cécile and Oskar Vogt Institute of Brain Research, University of Düsseldorf, Düsseldorf 40204, Germany 4 j.reckfort@fz-juelich.de

Description

Abstract: 3D-polarized light imaging (3D-PLI) reconstructs nerve fibers in histological brain sections by measuring their birefringence. This study investigates another effect caused by the optical anisotropy of brain tissue – diattenuation. Based on numerical and experimental studies and a complete analytical description of the optical system, the diattenuation was determined to be below 4 % in rat brain tissue. It was demonstrated that the diattenuation effect has negligible impact on the fiber orientations derived by 3D-PLI. The diattenuation signal, however, was found to highlight different anatomical structures that cannot be distinguished with current imaging techniques, which makes Diattenuation Imaging a promising extension to 3D-PLI.