Published October 23, 2026
| Version v1
Journal article
Open
Impact of ab initio line-intensity ratios on primary spectroscopic thermometry
Authors/Creators
Description
Line-intensity-ratio thermometry (LRT) is a spectroscopic primary thermometry method whose accuracy depends critically on the quality of the reference molecular line intensities. We investigate the impact of reference spectroscopic data on LRT by reanalysing previously measured cavity mode-dispersion spectroscopy spectra of ten CO (3–0) band transitions at 296 K using newly available ab initio line-intensity ratios [Science Advances 11, eadz6560 (2025)]. Keeping the experimental data and analysis procedure unchanged allows the contribution of the reference data to be isolated from other uncertainty sources. The updated line-intensity ratios improve the consistency of temperatures retrieved from different line pairs by an order of magnitude compared to the result reported in [Physical Review Letters 135, 103201 (2025)]; for seven out of nine line pairs, the statistical standard error of the mean (SEM) of retrieved temperatures is 1.3 mK. The weighted mean spectroscopic temperature agrees with the independently measured gas-cell temperature within a few mK. The combined standard uncertainty of the mean spectroscopic temperature is 19 mK (62 ppm at 296 K), with the dominant contribution arising from spectral line-shape modeling rather than reference line intensities.
Files
1-s2.0-S0263224126033026-main.pdf
Additional details
Related works
- Is published in
- Journal: 0263-2241 (ISSN)
Dates
- Available
-
2026-11-02