Published September 21, 2026 | Version v1

Method development and validation of Estimation of nitrofurantoin in tablet dosage form by RP-HPLC

Authors/Creators

Description

A rapid, sensitive, and selective method was developed and validated for the estimation of nitrofurantoin in tablet dosage forms using high-performance liquid chromatography (HPLC). Nitrofurantoin and acetanilide (internal standard) were analyzed using an Inertsil ODS-3 column (250 × 4.6 mm, 10 μm) with a mobile phase consisting of acetonitrile and phosphate buffer (6.8 g of potassium dihydrogen phosphate, adjusted to pH 7.0 with 1 N sodium hydroxide solution) in a ratio of 88:12 (v/v). The analytes were dissolved using a diluent consisting of dimethylformamide and water in a ratio of 40:50 (v/v). Detection was carried out using a photodiode array (PDA) detector. The retention times of nitrofurantoin and acetanilide (internal standard) were found to be 7.655 and 15.068 min, respectively. The calibration curve was linear over the concentration range of 282 to 836mg/ml with a correlation coefficient (r²) of 1.0000. The recovery was found to be 98.1%, 98.3%, and 98.6% at the 50%, 100%, and 150% levels, respectively. Furthermore, the method was found to be rugged and robust. The developed method is therefore suitable for the quantitative estimation of nitrofurantoin in tablet dosage forms

Files

1105413419.pdf

Files (263.1 kB)

Name Size
md5:03cdbdfca54c4ef86221b8c86382a032