Published September 28, 2026 | Version v1

Termometria 3D por Ultrassom

  • 1. Department of Physics, FFCLRP University of São Paulo (USP) Ribeirão Preto, Brazil
  • 2. Department of Physics, FFCLRP University of São Paulo (USP) / National Institute of Science and Technology in Innovative Research in Health Sciences: from Nanotechnology to Artificial Intelligence (INCT) Ribeirão Preto, Brazil

Description

Conventional thermal therapy monitoring often relies on high-cost modalities like MRI or is limited to 2D ultrasound planes, which may overlook critical offplane hotspots. This work presents the implementation of an automated theranostic platform designed for 3D volumetric hyperthermia monitoring. The system integrates an open-architecture ultrasound scanner (ArtUs EXT-1H) with a custom-built mechanical scanning module and a 927 kHz high-intensity focused ultrasound (HIFU) transducer. By applying the Thermal Strain Imaging (TSI) method and the Lucas-Kanade optical flow algorithm to raw radiofrequency (RF) data, the platform reconstructed the induced thermal field across a 1.62° angular sweep on a paraffin-gel phantom. Experimental results demonstrated a localized heating pattern consistent with the focal geometry, reaching a maximum temperature of 41.6 °C. The volumetric mapping allowed for the visualization of discrete thermal slices, providing a new path to analyze a discret thermal map volume. This establishes a robust framework for non-invasive therapeutic monitoring in public health settings.

Files

670 - Termometria 3D por Ultrassom.pdf

Files (423.4 kB)

Name Size
md5:81cf4cb1c10541a8209ddc31f0802ed4
423.4 kB Preview Download

Additional details

Related works

Is part of
2359-3164 (ISSN)