Non-invasive near-infrared estimation of blood concentration in chest tube output
Authors/Creators
Description
This record contains the dataset and analysis code supporting a study of non-invasive near-infrared measurement of blood concentration in chest tube output. Bovine blood was diluted in 0.9% sodium chloride at ten proportions from 0% to 90% blood in 10% increments, sealed in segments of chest tube, and measured five times at each level in randomized order, giving 50 measurements at six wavelengths from 660 to 940 nm. Each measurement is the mean of 2000 samples acquired at 100 Hz through the 10-bit analog-to-digital converter of an Arduino Nano, with a dark reading taken after each emitter activation and no filtering applied.
The accompanying Python script reproduces all reported results: the fitted calibration parameters, the inversion procedure by which an observed optical signal is assigned to a calibrated dilution level, apparent and cross-validated model performance under leave-one-tube-out and leave-one-measurement-out schemes, the Bland-Altman agreement analysis with cluster bootstrap confidence intervals, and the principal components analysis of the emitter set. Full column definitions and method notes are given in the README.