D2 Spectrum catalogue including spectra for reference radiation qualities currently described in IEC 61267 and the recommended new reference radiation qualities (RRQs)

  • 1. ROR icon Physikalisch-Technische Bundesanstalt
  • 2. ROR icon Český Metrologický Institut
  • 3. ROR icon VSL Dutch Metrology Institute
  • 4. ROR icon ENEA Casaccia Research Centre
  • 5. ROR icon Helsinki University Hospital
  • 6. ROR icon Radiation and Nuclear Safety Authority

Description

1        Spectrum catalogue

Spectra measured by the project partners were submitted to PTB for further evaluation. PTB reviewed the submitted data and calculated mean spectra based on the accepted submissions. Since the reported spectra were provided with differing energy bin structures, the data were rebinned prior to averaging using linear interpolation between the reported bins.

1.1       Data review

As part of the data review, spectra with dead times exceeding 13 % were excluded from the dataset. In addition, the project partners were asked to provide the reference distance used to establish the reported radiation qualities. To assess the influence of differing source-to-detector distances on the half-value layer (HVL), comparable reference spectra were used. The HVL of these spectra was first calculated without modification. Subsequently, the spectra were computationally filtered with an additional air path corresponding to the difference between the reported reference distance and the actual spectrometer measurement position, and the HVL was recalculated. If the resulting deviation in HVL exceeded 0.1 %, the corresponding submitted spectrum was excluded from further evaluation. The remaining accepted spectra were then used for the calculation of the mean spectra.

1.2       Content

The catalogue includes tables summarizing the corresponding characteristic quantities, including HVL, mean, energy and as well as related uncertainties based on the variance of the reported data from different laboratories. Calculations were performed according to a procedure described in literature [2]. In addition, figures were included to show the individual submissions.

 

Funder name: European Partnership on Metrology
• Funder ID: 10.13039/100019599
• Grant number: 22NRM01 TraMeXI

Files

data.zip

Files (1.2 MB)

Name Size Download all
md5:d59e01a1f34afc7ac34088950b039295
230.5 kB Preview Download
md5:94d545ee4c10937e023d6561917a1e15
694.8 kB Preview Download
md5:035f6c4692d4045f7da2367af07e7e31
307.9 kB Preview Download

Additional details

Funding

European Commission
TraMeXI 22NRM01