Published March 19, 2025
| Version v1
Conference paper
Open
Large-area dynamic contrast MHz optical coherence tomography for label-free imaging of porcine tissue
Authors/Creators
-
Burhan, Sazgar
(Researcher)1
-
Göb, Madita
(Researcher)1
-
Pieper, Mario
(Researcher)2
- Laedtke, Tjalfe (Researcher)1
- Grahl, Thorge (Researcher)1
- Münter, Michael (Researcher)1
-
Schulz-Hildebrandt, Hinnerk
(Researcher)3, 4
-
Hüttmann, Gereon
(Supervisor)1, 5
- König, Peter (Supervisor)1
-
Huber, Robert
(Supervisor)6, 1
Description
We demonstrate a 3.2 MHz-OCT system for inter-volumetric dynamic optical coherence tomography of ex vivo porcine kidney tissue. Employing a home-built Fourier Domain mode locking (FDML) laser with a 1310 nm wavelength, the system achieved a lateral resolution of 3.48 μm and a frame rate of 612 Hz. A motorized XYZ positioning stage enabled the precise acquisition of multiple volumes, which were seamlessly stitched together to generate a comprehensive dataset with a total area of 2.6 × 2.6 mm2. Validations against histological sections confirmed the system’s ability to visualize cellular tissue structures.
Files
BiOS_2025_SB_Zenodo.pdf
Files
(496.2 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:4f0c0fb7c0b7d8f07accdc9e6ef25d78
|
496.2 kB | Preview Download |