0.55T qualitative and quantitative knee MRI data for iHEALTH and A4IM projects

  • 1. Millenium Institute for Intelligent Healthcare Engineering
  • 2. Millennium Institute for Intelligent Healthcare Engineering

Description

Information for Users

Scanner

The MRI scanner used to acquire data is the MAGNETOM Free.Max 0.55T commercial system manufactured by Siemens Healthineers (Erlangen, Germany), FDA cleared in July 2021 (510(k) Number: K210611). The scanner was designed under the constraints of being compact, lightweight, cost-effective, and indicated for orthopedic practices and routine musculoskeletal imaging applications.

The scanner employs a superconducting magnet with DryCool technology to generate B₀ ≈ 0.55 T at the field of view. The magnet assembly weighs less than 3.5 metric tons with a system height of less than 80 inches. The bore diameter is 80 cm, optimized for patient comfort and accessibility, beneficial for patients with claustrophobia or anatomical challenges. The large bore also facilitates knee imaging with flexible coil positioning.

The gradient coil geometry employs actively shielded gradient technology with the following specifications: A maximum gradient amplitude of 22-26 mT/m, maximum slew rate of 45-55 T/m/s, a Field of View (FOV) up to 50 cm. The Tx/Rx RF coil used to acquire knee data is the Contour S Coil (Part Number: 11407176), a commercial 6-channel flexible receive coil manufactured by Siemens Healthineers (Erlangen/Forchheim, Germany). 

The Contour S Coil employs a blanket-like flexible design with a dimension of 45 cm × 27 cm (17.7 in × 10.6 in) that provides patient comfort by adapting to body contours without the need for straps. The coil offers ultra-lightweight construction with soft materials. The flexibility allows the coil to conform closely to the knee joint, maximizing signal-to-noise ratio (SNR) through improved coil-to-tissue coupling compared to rigid coil alternatives.

The coil operates at the proton Larmor frequency corresponding to the 0.55T field strength (≈23.6 MHz) and the 6-element phased array design enables parallel imaging acceleration for reduced scan times while maintaining diagnostic image quality.

Control system and data acquisition

The MR system is controlled through the manufacturer's proprietary software platform syngo MR XA60, which handles synchronized pulse-sequence execution and data acquisition. User interaction with the system is handled via the syngo MR user interface. The raw data is acquired in ISMRMRD format and the reconstructed images are available in DICOM format.

Subjects information

For this dataset, 20 subjects were recruited and both knees were scanned, resulting in a total of 40 knee images. Written informed consent was obtained from all volunteers prior to imaging. Anonymous information about the gender, age, height, and possible knee injuries of the volunteers scanned for this database is provided in Table 1.  The numbering of the subjects corresponds to the numbering specified in the folder and file names in this database.

Table 1. Additional anonymous information per subject.

Subject

Sex

Age (years)

Height (cm)

Weight (kg)

Comments

Subject 1

Male

68

177

97

Without injuries/anomalies in any knee

Subject 2

Male

34

172

72

Without injuries/anomalies in any knee

Subject 3

Male

47

183

94

Without injuries/anomalies in any knee

Subject 4

Female

30

157

58

Without injuries/anomalies in any knee

Subject 5

Male

28

176

64

Without injuries/anomalies in any knee

Subject 6

Female

60

159

69

Without injuries/anomalies in any knee

Subject 7

Male

24

173

96

Without injuries/anomalies in any knee

Subject 8

Male

30

169

85

Without injuries/anomalies in any knee

Subject 9

Male

35

170

77

Without injuries/anomalies in any knee

Subject 10

Male

28

169

83

Without injuries/anomalies in any knee

Subject 11

Female

27

164

57

Without injuries/anomalies in any knee

Subject 12

Male

31

174

73

Titanium Corkscrew suture anchors in the left knee

Subject 13

Male

27

165

70

Without injuries/anomalies in any knee

Subject 14

Female

38

164

70

Without injuries/anomalies in any knee

Subject 15

Male

57

182

100

Without injuries/anomalies in any knee

Subject 16

Male

53

175

90

Bone finding at the level of the femur in the right knee

Subject 17

Female

52

160

90

Without injuries/anomalies in any knee

Subject 18

Male

62

177

95

Bone finding at the level of the femur in the right knee

Subject 19

Male

52

173

95

Small meniscal tear in the left knee, meniscus surgery in the right knee, and chondromalacia

Subject 20

Male

33

186

120

Without injuries/anomalies in any knee

 

Pulse sequences

For each volunteer, the complete exam can be separated into two groups of sequences: a) a qualitative protocol and b) a quantitative protocol. 

The qualitative protocol consists of running three consecutive 2D sequences with cartesian trajectories, and optimized parameters to obtain Proton Density (PD), T1-weighted, and T2-weighted images. Deep Resolve deep learning vendor software image enhancement was applied to each sequence, resulting in a total of 6 datasets. The quantitative protocol consists of acquiring two sets of images: (1) 2D Modified Look-Locker inversion (MOLLI) (3-3-5) to obtain a T1 map, and (2) and  T2 prepared balance steady state free precession (T2p-bSSFP) sequence was acquired to obtain a T2 map. The 2D T1 and T2 maps were obtained with the vendor software package using DL for the T1 and T2 contrasts by fitting the 3-parameter model for MOLLI images and the T2 decay model for T2p-bSSFP images. The calculated maps and their corresponding image contrasts can be found in this dataset.

In all the cases the images were acquired  in the sagittal plane for better knee anatomic visualization. Each image set is acquired for the same subject in the same session for both knees, and the protocol requires approximately 8:42 minutes of sequence execution time per knee, giving a total time of 17:24 min per volunteer. 

The acquisition time per sequence performed for both types of protocols and per knee are listed in Table 2. The main acquisition parameters of the sequences performed for both protocols are listed in Tables 3 and 4. Example representative images for the qualitative and quantitative protocols are shown in Figure 1 and Figure 2.

Table 2. Acquisition time per sequence

Sequences

Time [min:sec]

AAKnee_Scout

0:32

pd_tse_sag_Right

0:57

t1_tse_sag_Right

0:39

t2_tse_sag_Right

0:44

pd_tse_sag_Right_DeepResolve

0:57

t1_tse_sag_Right_DeepResolve

0:39

t2_tse_sag_Right_DeepResolve

0:44

t1map_MOLLI_Right

1:24

t2map_trufi_Right

2:06

—Knee change—

—

AAKnee_Scout

0:32

pd_tse_sag_Left

0:57

t1_tse_sag_Left

0:39

t2_tse_sag_Left

0:44

pd_tse_sag_Left_DeepResolve

0:57

t1_tse_sag_Left_DeepResolve

0:39

t2_tse_sag_Left_DeepResolve

0:44

t1map_MOLLI_Left

1:24

t2map_trufi_Left

2:06

Total time

17:24 min

 

Table 3. Sequence parameters for quantitative protocol

 

PD-w

T1-w

T2-w

N° averages

1

1

1

FOV (cm)

17

17

17

Matrix size

112 x 112

112 x 112

112 x 112

Resolution (mm)

(0.8, 0.8, 5)

(0.8, 0.8, 5)

(0.8, 0.8, 5)

N° Slices

10 - 14

10 - 14

10 - 14

{Rd, Ph, Sl}

{x,y,z}

{x,y,z}

{x,y,z}

ETL

9

4

12

RF phase convention

CPMG

CPMG

CPMG

TE (ms)

28

12

78

k-Space trajectory

Cartesian

Cartesian

Cartesian

Inversion time (ms)

0

0

0

Repetition time (ms)

3070

490

3040

BW acq (Hz/Px)

100

120

110

Dummy pulses

0

0

0

Slice zero filling (%)

0

0

0

Tscan (min: sec)

0:57

0:39

0:44

 

Table 4. Sequence parameters for quantitative protocol

 

T1 Mapping

T2 Mapping

N° averages

1

1

FOV (cm)

20 x 20

20 x 20

Matrix size

128 x 128

128 x 128

Resolution (mm)

(1.6, 1.6, 8)

(1.6, 1.6, 8)

N° Slices

6

6

{Rd, Ph, Sl}

{x,y,z}

{x,y,z}

ETL

-

-

TE (ms)

2.08

1.98

k-Space trajectory

Cartesian

Cartesian

Inversion time (ms)

340

340

Repetition time (ms)

891.16

561.26

BW acq (Hz/Px)

543

558

Dummy pulses

8

5

Slice zero filling (%)

0

0

Tscan (min)

1:24

2:06

 

Figures can be found in the JPG files attached below:

Figure 1. Reference images for a) Proton density (PD) knee sagittal image, b) PD sagittal image with deep resolved, c) T1-weighted sagittal image, d) T1-weighted sagittal image with deep resolved, e) T2-weighted sagittal image, f) T2-weighted sagittal image with deep resolved. 

Figure 2. Reference images for a) knee sagittal 2D MOLLI T1 map, b) knee sagittal 2D T2p-bSSFP T2 map.

Files

Figure1.png

Files (14.8 GB)

Name Size
md5:dd5235e2a7c80315f10668010d531133
290.8 kB Preview Download
md5:ca7eafe1fdbb05d9f9399e198b478f53
236.6 kB Preview Download
md5:c1e2a0ef5901e4745fd7dd101e3475c2
723.4 MB Preview Download
md5:f1f7372df91214980697ebe3339f69ee
798.9 MB Preview Download
md5:1dfe43da0e2feaa8a739998002630005
652.1 MB Preview Download
md5:02a837427775e66d5c0770cca52522e5
754.7 MB Preview Download
md5:1fb0988c9e3503ba3cedc350fdb7e104
732.1 MB Preview Download
md5:547c27490f5262eccb9fee69c44d860e
725.0 MB Preview Download
md5:7df21e60ca9e2ce68a3697f36eb534fb
781.2 MB Preview Download
md5:f249543444299b4f41e474d47a7b8f92
796.3 MB Preview Download
md5:e8c3b824c291000831d78ef7814b765b
725.7 MB Preview Download
md5:ebdb85b37ca82e13769a2dce8a770705
800.5 MB Preview Download
md5:be30b2f5371fa50795bd28afc4ef7ab1
796.3 MB Preview Download
md5:dc2c1ed22ceb65ddc91212a58ea325b5
703.6 MB Preview Download
md5:9859c3ffb42efc77d1a8a7b8525a699c
699.2 MB Preview Download
md5:3145043d3f56591e4e7e669a75ecc0ef
725.0 MB Preview Download
md5:3f519ea57d67a758f0a72053adf38ce2
723.6 MB Preview Download
md5:5885909967cb1d6260f8cafa153f09a1
709.7 MB Preview Download
md5:3901f9fc3dde0d78edd0a91e95173048
710.7 MB Preview Download
md5:b7418dfcbcc18ec4df0daf4c77739344
863.2 MB Preview Download
md5:865a7f343c63fcb88de053826071a549
722.6 MB Preview Download
md5:7974de2b160e0c85b721007412bc58a2
702.8 MB Preview Download

Additional details

Funding

Agencia Nacional de Investigación y Desarrollo

Dates

Submitted
2026-03-26