Published October 3, 2025 | Version v1
Journal article Open

Traceability of S-Parameter Measurements Up to 167 GHz Using 0.8 mm Coaxial Standards

  • 1. ROR icon Physikalisch-Technische Bundesanstalt

Description

To establish traceable S-parameter measurements in the 0.8 mm coaxial connector system, dimensional measurements are performed on a set of seven o set short standards. Reflection coecients for each standard are calculated numerically and analytically based on these dimensions. S-parameter measurements are conducted for one-port devices up to 167 GHz, and
an optimization-based calibration is applied, treating material properties as free parameters. Selected measurement results are
presented, and associated uncertainty budgets are discussed.

Files

Traceability_of_S-Parameter_Measurements_Up_to_167_GHz_Using_0.8_mm_Coaxial_Standards.pdf

Additional details

Related works

Is identical to
Journal article: 10.1109/TMTT.2025.3613009 (DOI)

Funding

European Association of National Metrology Institutes
RF key quantities for 6G development 23IND03