Published October 3, 2025
| Version v1
Journal article
Open
Traceability of S-Parameter Measurements Up to 167 GHz Using 0.8 mm Coaxial Standards
Authors/Creators
Description
To establish traceable S-parameter measurements in the 0.8 mm coaxial connector system, dimensional measurements are performed on a set of seven o set short standards. Reflection coecients for each standard are calculated numerically and analytically based on these dimensions. S-parameter measurements are conducted for one-port devices up to 167 GHz, and
an optimization-based calibration is applied, treating material properties as free parameters. Selected measurement results are
presented, and associated uncertainty budgets are discussed.
Files
Traceability_of_S-Parameter_Measurements_Up_to_167_GHz_Using_0.8_mm_Coaxial_Standards.pdf
Files
(6.9 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:d0555e74832591e8b9f4bf78fe6f50ec
|
6.9 MB | Preview Download |
Additional details
Related works
- Is identical to
- Journal article: 10.1109/TMTT.2025.3613009 (DOI)