Optical behavior of an enhanced monofocal intraocular lens compared with a standard one
Authors/Creators
Description
Abstract: The aim of this work was to compare an enhanced monofocal (RayOne EMV RAO200E,
Rayner) and standard monofocal (RayOne RAO600C Aspheric, Rayner) intraocular lenses (IOLs)
for three nominal powers (+10.00 D, +20.00 D and +30.00 D) as a function of the optical aperture
diameter (pupil diameter) using a commercial Schlieren phase-shifting deflectometer NIMO TR1504
(Lambda-X, Belgium). From the wavefront maps measured by this instrument, the radial power
profiles, the spherical aberration coefficients of the Zernike polynomial expansion (as a function of
the optical aperture radius), and the root-mean-square (RMS) of the high-order aberrations (HOAs)
were obtained and analyzed by comparing the two models. The results showed that the effective
added power that could be obtained with the enhanced model depended directly on the pupil size
and the power of the IOL implanted. The higher additions were achieved with the higher nominal
IOL powers. The relationship between the pupil diameter, the corneal aberration of the patients and
the power profile of these IOLs could have a crucial implication on the far distance and the final
effective addition. However, it is important to note that these findings should be clinically validated
through the implantation of these models in patients’ lenses
Files
applsci-13-12938.pdf
Files
(2.0 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:5fbe8e42c6076b777de7ea0db91dc71f
|
2.0 MB | Preview Download |
Additional details
Funding
- Agencia Estatal de Investigación
- PID2021-122486OA-I00
Dates
- Available
-
2023-12-04
References
- García-Montero M, Garzón N, Gonzalez-Fernandez V, Gómez-Pedrero JA, Albarrán-Diego C. Optical Behavior of an Enhanced Monofocal Intraocular Lens Compared with a Standard One. Appl. Sci. 2023, 13, 12938. https://doi.org/10.3390/app132312938