Published June 19, 2025 | Version v1
Conference paper Open

Traceable S-Parameter Measurements up to 165 GHz using 0.8mm Coaxial Standards

  • 1. ROR icon Physikalisch-Technische Bundesanstalt

Description

To extend SI-traceable calibration for coaxial measurements up to 165 GHz, seven calculable o set short standards in the 0.8 mm coaxial system are used. The standards are
calculated based on dimensional measurements and used for an overdetermined least-squares calibration. Measurement results and uncertainty budgets are reported for a flush short and a broadband match. The repeatability of the interface and a minimal pin gap are investigated.

Notes

The project 23IND03 RF 4 6G has received funding from the European Partnership on Metrology, co-financed from the European Union’s Horizon Europe Research and Innovation Programme and by the Participating States.

• Funder name: European Partnership on Metrology
• Funder ID: 10.13039/100019599
• Grant number: 23IND03 RF 4 6G

Files

Schramm_Sparam_165GHz.pdf

Files (1.6 MB)

Name Size Download all
md5:1437295a8828cdd62d12e6044064afa5
1.6 MB Preview Download

Additional details

Funding

European Association of National Metrology Institutes
RF key quantities for 6G development 23IND03