Published June 19, 2025
| Version v1
Conference paper
Open
Traceable S-Parameter Measurements up to 165 GHz using 0.8mm Coaxial Standards
Authors/Creators
Description
To extend SI-traceable calibration for coaxial measurements up to 165 GHz, seven calculable o set short standards in the 0.8 mm coaxial system are used. The standards are
calculated based on dimensional measurements and used for an overdetermined least-squares calibration. Measurement results and uncertainty budgets are reported for a flush short and a broadband match. The repeatability of the interface and a minimal pin gap are investigated.
Notes
Files
Schramm_Sparam_165GHz.pdf
Files
(1.6 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:1437295a8828cdd62d12e6044064afa5
|
1.6 MB | Preview Download |