Published September 30, 2025
| Version v1
Report
Open
Report on results - the relationship between the power and the temperature difference for four implants
Contributors
Researchers:
Description
Dataset and final report investigting whether and to which extent it is possible to replace or support thermal measurements/simulations with power measurements/simulations for gradient heating testing of passive conductive implants according to the procedure defined in the ISO1094 standard.
Files
AnklePlate.zip
Files
(7.6 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:994c227349347704b622a7cfd0917031
|
1.2 MB | Preview Download |
|
md5:190a5fd710d7dba0b06bc1e0fccf9d05
|
2.8 MB | Preview Download |
|
md5:34fd08ee09965cfaf75fe90e101f39a2
|
1.2 MB | Preview Download |
|
md5:c0a969df92752b26ecbac7b4fa5941f8
|
1.2 MB | Preview Download |
|
md5:09319ff8e7a4f432dacbde17092092f8
|
70.1 kB | Preview Download |
|
md5:da9af8cc2c8854542a5eb68e39ff072f
|
1.2 MB | Preview Download |