Published June 24, 2008 | Version v1

Applying Spectroscopy to Global Change: Isotopic Ratio Measurements of Ambient Carbon Dioxide and Methane

  • 1. Aerodyne Research, 45 Manning Road, Billerica, MA, USA 01821

Description

Real time methods to monitor the stable isotopic ratios of carbon dioxide and methane are needed to quantify the sources and sinks of these important greenhouse gases. High resolution molecular spectroscopy provides an attractive solution. However, this is an extreme instrumental challenge since the ratios need to be measured with an accuracy of approximately one part in ten thousand or 0.1%. We use tunable infrared laser differential absorption spectroscopy with quantum cascade (QC) lasers to address this challenge. QC lasers are attractive optical sources since they do not need cryogenic fluids, they have stable single mode spectral output and offer optical simplicity. We have used a pulsed-QCL instrument to perform long term (>10 months) monitoring of ambient air from the roof top of our laboratory with time resolution of one second. The resulting data set shows both short term and long term isotopic variations. We observe clear seasonal changes in both 13C/12C and 18O/16O. We have recently extended this technique to the even more challenging case of 13CH4 using continuous wave QCLs. Preliminary results of these measurements will be presented as well.

Notes

Session IV-6.

Files

T4.6_Nelson_HITRAN_2008.pdf

Files (5.7 MB)

Name Size
md5:b0d144d9f65d7c802f1744e819b0d77d
5.7 MB Preview Download

Additional details

References

  • None given