Published October 21, 2025 | Version v1

Construction and Evaluation of Non-contact Fixed-point Cells

  • 1. EDMO icon University of Ljubljana, Faculty of Electrical Engineering

Description

Poster documents the adoption of radiation thermometer calibration capability using fixed points at ULFE/LMK. Four miniature fixed-point cells for pyrometric use have been filled and tested in scope of the MultiFixRad project, namely: copper, iron-carbon, cobalt-carbon and palladium-carbon eutectic alloy cells. Data processing for melting/freezing temperature classification is conducted using smoothened numerical first- and second-order derivatives of measurement timeseries.

Files

poster_MFR_VM.pdf

Files (2.3 MB)

Name Size Download all
md5:530f5bc3e2d50781e6c6f062391cb172
2.3 MB Preview Download

Additional details

Funding

European Metrology Programme for Innovation and Research
MultiFixRad 22RPT03