Published September 12, 2024 | Version v1
Dataset Open

Drift of Pt-Rh thermocouples in emf and in temperature terms at 1324 °C

  • 1. ROR icon National Physical Laboratory

Description

Dataset containing drift of selected Pt-Rh thermocouples which are part of a multi-wire thermocouple. Drift is in emf terms and in temperature terms. Drift was measured by exposing the thermocouples to a temperature of approximately 1324 °C, with periodic in-situ calibration using a Co-C 'fixed-point' point cell (melting point 1324.29 °C).

Files

Files (21.5 kB)

Name Size Download all
md5:fed0a5f313740efa38dcb27b28a9cca7
21.5 kB Download

Additional details

Funding

UK Research and Innovation
Fundamental principles of sensor network metrology (FunSNM) 10097635