There is a newer version of the record available.

Published January 8, 2018 | Version v2

Aethalometer data processor

Authors/Creators

Description

Data acquired from filter-based measurement such as Aethalomenter needs carefully correction due to its inherent systemic error, i.e., filter matrix effect, scattering effect and loading effect. The non-linear relationship between  (light attenuation on the filter) and  (light absorption coefficient of suspended aerosol in ambient air) has become a challenge for Aethalomenter, as a linear relationship is assumed by default in its output data. This program applies Weingartner (2003) and Virkkula (2007) algorithm to correct the data. QA/QC features are also provided, including statistic of sensor voltage.

 

Please cite this paper if Aethalometer data processor is used in your publication:

 

Wu, C., Wu, D., and Yu, J. Z.: Quantifying black carbon light absorption enhancement with a novel statistical approach, Atmos. Chem. Phys., 18, 289-309, doi:10.5194/acp-18-289-2018, 2018.

 

The latest version of the program can be found at:

https://sites.google.com/site/wuchengust/

https://doi.org/10.5281/zenodo.832403

 

 

================================================================================================

 

从基于膜采样的测量获得的吸收系数数据(例如Aethalomenter)需要仔细校正,因为其固有的系统误差,包含采样膜的matrix效应,散射效应和载荷效应。 (采样膜上的光衰减)和 (环境空气中悬浮的气溶胶的光吸收系数)之间的非线性关系已成为Aethalomenter的挑战,因为在其输出数据中默认地假设为线性关系。 该程序应用Weingartner(2003)和Virkkula(2007)算法来校正数据。 还提供QA / QC特性,包括传感器电压的统计等信息。

 

如果发表的文章中使用了本程序处理数据,请引用以下文章:

Wu, C., Wu, D., and Yu, J. Z.: Quantifying black carbon light absorption enhancement with a novel statistical approach, Atmos. Chem. Phys., 18, 289-309, doi:10.5194/acp-18-289-2018, 2018.

 

最新版本的程序可以在以下网站找到:

https://sites.google.com/site/wuchengust/

https://doi.org/10.5281/zenodo.832403

 

 

Files

AEprocessor_20170315.zip

Files (5.5 MB)

Name Size
md5:318940641f91329b0d585afaffdc3d29
2.5 MB Preview Download
md5:96273c0e65febe3808169019adc346d0
3.0 MB Preview Download