Published January 15, 2018
| Version v1
Journal article
Open
Diode laser-based gas analyser for the simultaneous measurement of CO<sub>2</sub> and HF in volcanic plumes
Authors/Creators
- 1. INGV, National Institute for Geophysics and Vulcanology, Pisa, Italy
- 2. CNR – National Institute of Optics, Firenze, Italy
- 3. University of Manchester, School of Earth and Environmental Science, Manchester, UK
Description
A portable analyser is described for simultaneous detection of CO2 and HF
emitted by volcanoes and fumaroles. The system is based on two
fibre-coupled distributed feedback lasers and only one multipass cell, and
provides the absolute concentration values of the two gases, without
requiring a calibration procedure, at a maximum rate of 4 Hz. The
spectrometer can operate both in a closed-cell configuration and in an
open-cell set-up, with the latter mitigating problems associated with
chemisorption of the HF molecule. The concept, practical realization and
laboratory performance of the device are presented, together with results
from a first test campaign measuring volcanic gases emitted from the crater of
Vulcano, Italy. We obtained an in-field sensitivity of 320 ppb for CO2 and
20 ppb for HF at 2 s integration time.