Published April 10, 2017 | Version version from 2017-11-12

Measurements of PM10 ions and trace gases with the online system MARGA at the research station Melpitz in Germany . A five-year study

  • 1. Atmospheric Chemistry Department (ACD), Leibniz Institute for Tropospheric Research (TROPOS)
  • 2. Im Dammwald 21, 61381 Friedrichsdorf, Germany
  • 3. Research Group for Environmental Monitoring, Alte Jonastraße 83, 8640 Rapperswil, Switzerland
  • 4. German Federal Environment Agency, Wörlitzer Platz 1, 06844 Dessau-Roßlau, Germany

Description

Abstract:  An hourly quantification of inorganic water-soluble PM10 ions and corresponding

trace gases was performed using the Monitor for AeRosols and Gases in ambient Air

(MARGA) at the TROPOS research site in Melpitz, Germany. The data availability amounts

to over 80% for the five-year measurement period from 2010 to 2014. Comparisons were

performed for the evaluation of the MARGA, resulting in coefficients of determinations

(slopes) of 0.91 (0.90) for the measurements against the SO2 gas monitor, 0.84 (0.88), 0.79

(1.39), 0.85 (1.20) for the ACSM NO3-. SO42- and NH4+ measurements respectively,

and 0.85 (0.65), 0.88 (0.68), 0.91 (0.83), 0.86 (0.82) for the filter measurements of Cl−, NO3−,

SO22− and NH4+, respectively.

 

 