Published October 12, 2023 | Version v1
Journal article Open

A VALIDATED REVERSE PHASE-HPLC-PDA METHOD AND OPTIMIZATION OF METHOD AND ITS VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF SULFADOXINE AND PYRIMETHAMINE IN PURE AND PHARMACEUTICAL DOSAGE FORM

Description

A new, simple and accurate, precise RP-HPLC method was developed for simultaneous determination of Sulfadoxine and Pyrimethamine in bulk and in combined pharmaceutical dosage form. The separation of Sulfadoxine and Pyrimethamine was achieved within 8 minutes on an Agilent Zorbax (C18) (150mm x 4.6mm, 5µm) column using Methanol: Acetate Buffer pH-3.8 (24:76v/v) as the mobile phase. Detection was carried out using wavelength at 262nm. The method showed adequate sensitivity concerning linearity, accuracy and precision over the range 100-500μg/ml and 30-70μg/ml for Sulfadoxine and Pyrimethamine, respectively. Careful validation proved advantages of high sensitivity, accuracy, precision, selectivity, robust and suitability for quality control laboratories. The developed method was robust as the %RSD was within the range and without effecting system suitability parameters. The proposed method is suitable for simultaneous determination of Sulfadoxine and Pyrimethamine in bulk and pharmaceutical dosage form.

Keywords: Sulfadoxine and Pyrimethamine, RP-HPLC, Validation, Precision, Robustness.