Published January 24, 2011 | Version v1
Journal article Open

Detection of Chemical Vapors Using Oscillator with Surface Acoustic Wave

Description

Abstract and figures

Surface Acoustic Wave (SAW) sensors demonstrate superior sensitivity in detection of chemical agents. Due to their solid state design and fabrication compatible with modern technologies, SAW chemical sensors are extremely reliable. The sensor is modeled as a two-port device with parts represented by equivalent circuits. Change of output frequency as a function of vapor concentration is calculated. The oscillator circuit with SAW sensor in the feedback loop is designed. The SAW sensor is used for modifying the oscillator frequency. The presence of chemical vapor is then detected by monitoring this frequency shift.

Files

Detection_of_Chemical_Vapors_Using_Oscillator_with.pdf

Files (454.8 kB)