Published May 24, 2023 | Version v2

Open-Source System for Real Time Functional Assessment of In-Vitro Filtration Barriers

  • 1. Icahn School of Medicine at Mount Sinai

Description

Microphysiological systems that incorporate synthetic or natural membranes with human cells mimic the human kidney, and they can readily be used to study key cell biological processes as well as drug effects and interactions. We present an affordable, open-source platform for the real-time monitoring of functional filtration status in engineered microphysiological systems. Using readily available components, our assay can linearly detect concentrations of two target molecules, FITC-labeled inulin and Texas Red-labeled human serum albumin, within clinically relevant ranges, and can be easily modified to integrate with different target molecules of different sizes and tags. We show that the platform enables real-time tracking of size-dependent diffusion with minimal fluid volume loss and without manual extraction of media, making it suitable for continuous operational monitoring of filtration status of microphysiological systems. The platform’s affordability and integrability with microphysiological systems make it ideal for many precision medicine applications, including evaluation of drug nephrotoxicity assessment and other forms of drug discovery applications.

Notes

Dataset includes raw intensity values and images in .TIFF format. Code can be found at: https://github.com/AzelogluLab/Sensor/

Files

Experimental Data.zip

Files (1.2 GB)

Name Size
md5:7a2ddda9e4d5d9c23e17c824f2b19a35
691.6 MB Preview Download
md5:13b477fe4252e3f1ed94377785de8d17
469.0 MB Preview Download