Published November 30, 2022
| Version v1
Project deliverable
Open
QuantumPascal D2 Report evaluating the performance of the different types of FP-based refractometers developed with respect to their precision, accuracy, working range and target relative uncertainties of 500 ppm in the range 1 Pa – 1 kPa and 10 ppm in the range 1 kPa – 100 kPa
Authors/Creators
- 1. Measurement Science and Technology, RISE Research Institutes of Sweden, SE-501 15 Borås, Sweden
- 2. Department of Physics, Umeå University, SE-901 87 Umeå, Sweden
- 3. Conservatoire national des arts et métiers (Cnam), LNE-Cnam, 61 rue du Landy, Saint-Denis, France
- 4. Physikalisch-Technische Bundesanstalt (PTB), Abbestr 2-12, Berlin, Germany
- 5. Centro Español de Metrología: Tres Cantos, Madrid, Spain
Description
This report summarize performance of the Fabry-Perot refractometers developed in work package 1 in the 18SIB04 QuantumPascal project. The report first briefly introduces the concept of performance in terms of precision, accuracy, and working range. Thereafter each of the systems that have been developed by the partners is shortly described and their contribution to the overall performance in terms of uncertainty is highlighted. From the combined results of the systems, it can be concluded that the target uncertainties have been reached.