Electromagnetic and RF Pulse Design Simulation Based Optimisation of an Eight Channel Loop Array for 11.7 T Brain Imaging

  • 1. Imaging Centre of Excellence, University of Glasgow, Glasgow, United Kingdom
  • 2. University of Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif sur Yvette, France
  • 3. Siemens Healthcare SAS, Saint-Denis, France

Description

Purpose: Optimisation of transmit array performance is crucial in ultra-high-field MRI scanners such as 11.7 T due to the increased radiofrequency (RF) losses and RF nonuniformity. This work presents a new workflow to investigate and minimise RF coil losses, and to choose the optimum coil configuration for imaging.


Methods: An 8-channel transceiver loop-array was simulated to analyse its loss mechanism at 499.415 MHz. A folded-end RF shield was developed to limit radiation loss and improve the B1+ efficiency. The coil element length, and the shield diameter and length were further optimised using electromagnetic (EM) simulations. The generated EM fields were utilised to perform RF pulse design (RFPD) simulations under realistic constraints. The chosen coil design was constructed to demonstrate performance equivalence in bench and scanner measurements.


Results: The use of conventional RF shields at 11.7 T resulted in significantly high radiation losses of 18.4%. Folding the ends of the RF shield combined with optimising the shield diameter and length increased the absorbed power in biological tissue and reduced the radiation loss to 2.4%. The peak B1+ of the optimal array was 42% more than the reference array. Phantom measurements validated the numerical simulations with a close match of within 4% of the predicted B1+.


Conclusion: A workflow, which combines EM and RFPD simulations, to numerically optimise transmit arrays was developed. Results have been validated using phantom measurements. Our findings demonstrate the need for optimizing the RF shield in conjunction with array element design to achieve efficient excitation at 11.7 T.