Published October 31, 2022
| Version v1
Project deliverable
Open
NFFA-Europe Pilot - D12.2 - Prototype of corrected CRL nanofocusing unit for user experiments
Description
We report on the implementation of a demonstrator for aberration correction of X-ray nanobeams generated by compound refractive lenses (CRLs) at beamline P06 of PETRA III at DESY. The aberrations are compensated by a tailor-made optical phase plate that needs to be positioned with respect to the CRLs in order to effectively compensate shape errors. The demonstrator combines the CRL mount and phase plate mount into a single device, allowing for a combined alignment. A novel piezo-driven positioning system allows for a one-off alignment of the phase plate relative to the CRL and freeze the alignment. In subsequent uses, the device can be operated by non-expert users.