Published January 30, 2023 | Version v1

Commissioning of the Iseult CEA 11.7T Whole-Body MRI: current status, gradient-magnet interaction tests and first imaging experience

Description

Objectives
The Iseult MRI is an actively shielded whole-body magnet providing a homogeneous and stable magnetic field of 11.7 T. After nearly 20 years of research and development, the magnet successfully reached its target field strength for the first time in 2019. This article reviews its commissioning status, the gradient–magnet interaction test results and first imaging experience.


Materials and Methods
Vibration, acoustics, power deposition in the He bath, and field monitoring measurements were carried out. Magnet safety system was tested against outer magnetic perturbations, and calibrated to define a safe operation of the gradient coil. First measurements using parallel transmission were also performed on an ex-vivo brain to mitigate the RF field inhomogeneity effect.


Results
Acoustics measurements show promising results with sound pressure levels slightly above the enforced limits only at certain frequency intervals. Vibrations of the gradient coil revealed a linear trend with the B0 field only in the worst case. Field monitoring revealed some resonances at some frequencies that are still under investigation.


Discussion
Gradient-magnet interaction tests at up to 11.7T are concluded. The scanner is now kept permanently at field and the final calibrations are on-going to pave the road towards the first acquisitions on volunteers.

Files

CleanManuscriptVersion.pdf

Files (1.8 MB)

Name Size
md5:e46a6be6bb64e8ba1153c9572bc48561
1.8 MB Preview Download

Additional details

Funding

European Commission
AROMA - Accurate, Reliable and Optimized functional MAgnetic resonance imaging at unprecedented field strength for unique exploration of the human brain 885876