Equipment Validation of Moist Heat Sterilization
Authors/Creators
Description
The purpose of this paper is to provide an overview of the process validation and control validation protocols of a steam sterilizer for an injectable room in a pharmaceutical laboratory. The process validation is carried out by using a series of tests, including a Bowie Dick test, a vacuum leak test, and a swab test. This test is conducted to confirm that the steam is adequately penetrating into the deepest regions of the load being steered. The temperature variation between 1210C and 1240C during the steering hold time suggests that the load has no impact on the uniform heating procedure that was obtained in the empty chamber heat distribution investigation.