Deliverable D7: "Validation report on the extended prediction model for multi-infusions."
Authors/Creators
- 1. METAS
- 2. NEL
- 3. UMCU
- 4. THL
Description
Multi-infusion setups are often used in critical care units such as the Intensive Care Unit (ICU) or the Operation Room (OR), where multiple medications are delivered through one single lumen of a thin catheter. The pharmacokinetics of the medications as well as the physical behaviour and effects of the multi-infusion setups are important aspects to understand the delivery of mixtures of several medications. A predictive model for dosing errors including the effect of the length of the catheter, the mechanical compliance of the plastic syringe and the flow inside the lines exhibiting a Poiseuille flow profile has already been reported [1]. This existing model has been extended to include the effect of air bubbles in infusion lines with and without an air filter and the presence of a non-return valve in infusion lines. Several experimental validation measurements have been performed and are described here in this report to validate the extended predictive model for dosing errors of multi-infusion setups.