Capillary glass viscometers are widely used and their method is well established worldwide for traceable measurement of the kinematic viscosity to the SI units. However, this technique does not allow in-line measurement of the dynamic viscosity in flow applications, which is important for the process-oriented information about the dynamic viscosity of the liquid. The micro pipe viscometers enable the characterization of liquids under flow conditions and the direct calibration of devices that measure dynamic viscosity under flow conditions or investigation of the influence of dynamic viscosity on the performance of flow devices.

This report describes the primary standards developed by NEL, RISE and METAS based on micro pipe viscometers for the in-line measurement of the dynamic viscosities. Their stated measurement uncertainties have been validated by the measurements of the dynamic viscosity of the liquids A – H and their consistency has been demonstrated in this report.