On the Feasibility of Using Behavioral Listening Effort Test Methods to Evaluate Cochlear Implant Fitting

  • 1. Department of Otorhinolaryngology and Head and Neck Surgery, Erasmus Medical Center, Rotterdam, The Netherlands

Description

Objective: Realistic outcome measures that reflect everyday hearing challenges are needed for evaluating hearing aid (HA) and cochlear implant (CI) fitting. According to literature, measures of listening effort may be more sensitive to differences between hearing-device settings than established measures of speech intelligibility. However, currently it is not clear which test is best suited to measure listening effort. The aim of the current study is therefore to investigate the feasibility of two listening effort tests to measure changes in listening effort due to differences in signal-to-noise ratio (SNR), as would arise from different hearing-device settings.

Design: Two behavioral listening effort tests were performed at two SNRs where the intelligibility was high: +4 dB and +8 dB above the 50% speech perception threshold. The authors compared a sentence final word identification and recall test (SWIRT), and a sentence verification test (SVT) in a group of 19 Dutch CI users on two occasions: test and retest. SWIRT measured the ability to recognize and recall the final words of sentences after a list of five or seven sentences was presented. SVT measured the ability and reaction time to determine whether a sentence was true or false. A better final-word recall in SWIRT and faster classification in SVT are believed to be related to less listening effort. Moreover, the final-word recognition in SWIRT and classification ability in SVT were recorded as measures of speech intelligibility. The authors related the sensitivity of the listening effort measures to the SNR differences to their accuracy by calculating test-retest differences. With a comparison to the speech intelligibility measures the authors evaluated whether these listening effort measures are better suitable to evaluate CI fitting.

Results: Results showed a learning effect and some outlying test items for SVT, but not for SWIRT. Both SVT and SWIRT could measure a significant difference in listening effort between the two SNR conditions, but the sensitivity of the listening effort measures to the SNR differences was low relative to their accuracy. SWIRT also measured a significant difference in speech intelligibility between the two SNR conditions, which had the highest sensitivity compared to its accuracy.

Conclusions: Of both tests, SWIRT is most suitable for measuring a difference in listening effort at group level due to an SNR difference in CI users. However, against the authors’ expectations, speech intelligibility scores are more suitable than the listening effort measures from SVT and SWIRT to evaluate and compare hearing-device settings in CI users.