Initial Experimental Validation of a Microwave Imaging System to Monitor Liver Microwave Thermal Ablation

Description

This study presents the initial experimental

validation of a mono-static microwave imaging system to

monitor liver microwave thermal ablation. The system consists

of a tank filled with a coupling medium in which a 3D-printed

phantom mimicking the ablation region is embedded. A slotloaded

Vivaldi antenna is moved in front of the phantom along

a linear path, measuring the S-parameters on a finite number of

positions. Both simulation and measurement results are found

to agree each other. The differential signal of the microwave

imaging system with and without the presence of the phantom

is over -90 dB, which is above the detection capacity of

commercial VNA. The experimental validation of the system

paves the way for the design of a multi-static microwave

imaging system.