Published March 2, 2022
| Version v1
Journal article
Open
Locally Engineered Detachment System Rescuing Aneurysm Coiling when Regular Balt Coil Release Device Used to Detach Balt Coils Had a Technical Error
Authors/Creators
- 1. Department of Neurosurgery, Annapurna Neurological Institute and Allied Sciences
- 2. Department of Biomedical Engineering, Annapurna Neurological Institute and Allied Sciences
- 3. Department of Radiology, Annapurna Neurological Institute and Allied Sciences
Description
Abstract :
While performing posterior communicating aneurysm coiling, the framing of the aneurysm was performed but the coil detachment device was having an error which was only known when the coil was released into the aneurysm. The biomedical engineer within the operating room created a twelve voltage power supply circuit which was used to detach all the coils, thus rescuing the situation and saving the patient.