Factors Affecting the Visceral Suitability of a Multibranched off-the-Shelf Endograft for the Treatment of Thoraco-Abdominal Aortic Aneurysms.

  • 1. Operative Unit of Vascular Surgery, IRCCS Policlinico San Donato, San Donato Milanese, Milan, Italy. Electronic address: daniela.mazzaccaro@gmail.com.
  • 2. Operative Unit of Vascular Surgery, IRCCS Policlinico San Donato, San Donato Milanese, Milan, Italy.
  • 3. Operative Unit of Vascular Surgery, IRCCS Policlinico San Donato, San Donato Milanese, Milan, Italy; Department of Biomedical Sciences for Health, University of Milan, Milan, Italy.

Description

Dataset from Mazzaccaro D, Avishay D, Nano G. Factors Affecting the Visceral Suitability of a Multibranched off-the-Shelf Endograft for the Treatment of Thoraco-Abdominal Aortic Aneurysms. Ann Vasc Surg. 2021 Nov 4:S0890-5096(21)00790-1. doi: 10.1016/j.avsg.2021.08.053. Epub ahead of print. PMID: 34743010.

Abstract

Background: To assess the factors affecting visceral suitability of the use of the Zenith T-branchTM system in a group of patients with thoraco-adbominal aortic aneurysms (TAAAs).

Methods: Computer tomography angiography (CTA) of patients who presented a TAAA from 01/2015 to 12/2019 were retrospectively examined. Multi-Planar Reconstructions were performed on CTA images to assess the anatomic suitability of the Zenith T-branch in the visceral district. In particular, the branch deviation angle (BDA), and the branch-length were computed for each target vessel.

Results: Fifty-four CTA were examined. In 33.3% of these patients the presence of either a common origin of the superior mesenteric artery and the celiac trunk, or the diameter of 1 or more visceral/renal artery limited the visceral suitability of the device. All patients except 1 (97.9%) fitted the BDA criterion when the graft was placed in a position in which the BDA for the SMA was 5 degrees to the left. The branch-length criteria was met in all patients, except for 1 (97.9%), when the graft was placed in the center of the aorta. The eccentrical placement of the endograft decreased the suitability to 93.7%.

Conclusions: The Zenith T-branch system can be suitable in the visceral district for about 67% of patients. The target artery diameter was the most limiting criterion. The central location of the graft within the aortic lumen significantly affected the branch-length distance criteria.

Files

Restricted

The record is publicly accessible, but files are restricted. Log in to check if you have access.

Request access

If you would like to request access to these files, please fill out the form below.

You need to satisfy these conditions in order for this request to be accepted:

These data are available upon duly motivated request to the corresponding author of the article

You are currently not logged in. Do you have an account? Log in here