Outcome of transcatheter aortic valve replacement in bicuspid aortic valve stenosis with new-generation devices.
- 1. Department of Clinical and Interventional Cardiology, IRCCS Policlinico San Donato, San Donato Milanese, Italy.
- 2. Department of Integrated Surgical and Diagnostic Sciences, University of Genoa, Genoa, Italy.
- 3. Department of Electrical, Computer, and Biomedical Engineering, University of Pavia, Pavia, Italy.
- 4. Department of Clinical and Interventional Cardiology, Cardio Center, Humanitas Research Hospital, Rozzano-Milano, Italy.
- 5. Department of Clinical and Interventional Cardiology, Humanitas Gavazzeni Hospital, Bergamo, Italy.
- 6. Department of Civil Engineering and Architecture, University of Pavia, Pavia, Italy.
Description
Dataset from Gorla R, Casenghi M, Finotello A, De Marco F, Morganti S, Regazzoli D, Bianchi G, Acerbi E, Popolo Rubbio A, Brambilla N, Testa L, Castriota F, Auricchio F, Reimers B, Bedogni F. Outcome of transcatheter aortic valve replacement in bicuspid aortic valve stenosis with new-generation devices. Interact Cardiovasc Thorac Surg. 2021 Jan 1;32(1):20-28. doi: 10.1093/icvts/ivaa231. PMID: 33201993.
Abstract
Objectives: To compare device success and paravalvular leak rates of 3 new-generation transcatheter aortic valve replacement devices in patients with bicuspid aortic valve stenosis and to test their biomechanical performance in a computer-based simulation model of aortic root with increasing ellipticity.
Methods: This retrospective multicentre study included 56 bicuspid aortic valve patients undergoing transcatheter aortic valve replacement with new-generation devices: Lotus/Lotus Edge (N = 15; 27%), Evolut-R (N = 20; 36%) and ACURATE neo (N = 21; 37%). Three virtual simulation models of aortic root with increasing index of eccentricity (0-0.25-0.5) were implemented. Stress distribution, stent-root contact area and paravalvular orifice area were computed.
Results: Device success was achieved in 43/56 patients (77%) with comparable rates among Lotus (87%), Evolut-R (60%) and ACURATE neo (86%; P = 0.085). Moderate paravalvular leak rate was significantly lower in the Lotus group as compared to Evolut-R group (0% vs 30%; P = 0.027) and comparable to the ACURATE neo group (0% vs 10%; P = 0.33). By index of eccentricity = 0.5, Lotus showed a uniform and symmetric pattern of stress distribution with absent paravalvular orifice area, ACURATE neo showed a mild asymmetry with small paravalvular orifice area (1.1 mm2), whereas a severely asymmetric pattern was evident with Evolut-R, resulting in a large paravalvular orifice area (12.0 mm2).
Conclusions: Transcatheter aortic valve replacement in bicuspid aortic valve patients with new-generation devices showed comparable device success rates. Lotus showed moderate paravalvular leak rate comparable to that of ACURATE neo and significantly lower than Evolut-R. On simulation, Lotus and ACURATE neo showed optimal adaptability to elliptic anatomies as compared to Evolut-R.