Published November 30, 2006 | Version v1

Hydride generation technique in flame atomic absorption spectrometry Determination of antimony in the presence of thorium matrix

Description

Radio chemistry Division, Bhabha Atomic Research Centre. Trombay, Mumbai-400 085, India

E-mail : svgod@barc.gov.in

Manuscript received 7 August 2006, accepted 21 August 2006

A method has been developed for the determination of Sb in the presence of thorium matrix, using hydride generation technique in Flame Atomic Absorption Spectrometry with air-acetylene as oxidant fuel combination.Studies have been carried out in detail to obtain the maximum absorbance signal due to the formation of Sb hydride in presence of thorium matrix. The efficiency of which is normally governed by the factors such as oxidant-fuel flow rate,KI concentration, hydrochloric acid molarity and concentration of thorium matrix. This has led to the determination of antimony to a level as low as 0.4 pm on thorium basis using 5 mg/ml Th matrix.Analysis of synthetic samples revealed a good agreement between the estimates obtained with those of added amounts with a precision better than 5% RSD. It is of significance to note that the present method gains its uniqueness for its ability to determine the analyte without the need for a prior chemical separation of a matrix as complex as Th.

Files

1135-1138.pdf

Files (424.1 kB)

Name Size
md5:58d04fa5d6d1238750058586fcbdc1ce
424.1 kB Preview Download