A highly selective and sensitive plasticized membrane sensor for holmium(lll) ions, based on 2-acetylfurane thiosemicarbazone (AFT) as membrane carrier, was prepared. The sensor shows a linear dynamic range of 1.0 x 10-6 to 1.0 x 10-2 M, with a Nernstian slope of 19.8 ± 0.3 mV per decade, and a detection limit of 5.5 x 10-7 M. It has a fast response time of ~ 5 s and could be used in a pH range of 2.3-7 .8. The proposed sensor revealed very good selectivities with respect to most of the common metal ions including Yb, Sm, Pr, Gd, Er, Tm, Cr, Fe, Na, K, Ca, Cd, Ni and Pb ions. The practical utility of the sensor has been demonstrated by .using it as an indicator electrode in the potentiometric titration of HoIII with EDT A, determination of fluoride ions in two mouth wash preparations and also to the determination of HoIII ions in tap water and river water samples.