Published October 3, 2021 | Version v1

Calibration of a Digital Current Transformer Measuring Bridge: Metrological Challenges and Uncertainty Contributions

  • 1. METAS

Description

In this paper, we consider the calibration of measuring bridges for non-conventional instrument transformers with digital output. In this context, the main challenge is represented by the necessity of synchronization between analog and digital outputs. To this end, we propose a measurement setup that allows for monitoring and quantifying the main quantities of interest. A possible laboratory implementation is presented and the main sources of uncertainty are discussed. From a metrological point of view, technical specifications and statistical analysis are employed to draw up a rigorous uncertainty budget of the calibration setup. An experimental validation is also provided through the thorough characterization of the measurement accuracy of a commercial device in use at METAS laboratories. The proposed analysis proves how the calibration of measuring bridges for non-conventional instrument transformers requires ad hoc measurement setups and identifies possible space for improvement, particularly in terms of outputs’ synchronization and flexibility of the generation process.

Files

METAS_Calibration_MeasuringBridge.pdf

Files (968.9 kB)

Name Size
md5:5aab713d0970982632b5fbe5baf72840
968.9 kB Preview Download