LIBS for real-time monitoring of laser-induced thermal damage in laserosteotomy: from dehydration to full carbonization
Authors/Creators
- 1. University of Basel
Description
Laserosteotomes will only be clinically applied if they do not carbonize the tissue. This canonly be ensured with an irrigation system with correctly tuned water and air flow rate. Well-designed irrigation helps avoiding thermal damage, increases ablation speed, andaccelerates healing while too little water leads to bone carbonization, and too much waterdecreases the ablation efficiency [1]. State-of-the-art for determining the irrigationparameters is to apply water and air flow rates based on pre-performed experimentswithout monitoring the bone surface in real-time. The lack of real-time monitoring of thelaser-induced thermal damage can thus result in carbonization, caused by either amalfunction in the irrigation system or by unexpected changes in the properties of thetissues from patient-to-patient (or even point-to-point due to an inhomogeneousstructure). Authors recently demonstrated that LIBS can classify fully carbonized bone fromnot carbonized one [2]. While such a classification can avoid further carbonization, evenlocalized carbonization on the surface of the bone can still prolong the healing process.There are several stages before full carbonization occurs, such as dehydration and earlycarbonization. The detection of these early signs makes prospective carbonizationprediction possible. This study aims at finding a regression between all steps of laser-induced thermal damage. Therefore, the spectrum of localized plasma generated in theablation spot, together with the number of shots sent to the bone without irrigation wasrecorded using an in-house fiber-coupled Echelle spectrometer. Later on, based on therecorded spectra, regression models were trained. The model used were regression tree(single tree and also an ensemble of several trees), support vector machine, and aconvolutional neural network. Finally, we calculated the root mean square error of eachmodel and applied the trained models to the data collected afterwards in real-time.
References:
[1]H. Abbasi, et al., “Effect of cooling water on ablation in Er:YAG laserosteotome of hardbone”, Proc. SPIE 10453 (2017).
[2] H. Abbasi, et al., “Laser-induced breakdown spectroscopy as a potential tool for autocarbonization detection in laserosteotomy”, J. Biomed. Opt. 23 (2018).
Acknowledgement:
The authors gratefully acknowledge funding of the Werner Siemens Foundation through the Minimally Invasive Robot- Assisted Computer-guided LaserosteotomE (MIRACLE) project.