BAM reference data: XPS raw data of Al-coated titania nanoparticles (JRCNM62001a and JRCNM62002a)

Radnik, Jörg

The raw data are given as VAMAS-File. The measurement condtions are given in the file. The C1s-fits are provided as ascii-files.

For further information please look at Radnik, J. Kersting, R., Hagenhoff, B., Bennet, F., Ciornii, D.; Nymark, P., Grafström R. and Hodoroaba, V.-D. Nanomaterials 2021, 11, 639. https://doi.org/10.3390/nano11030639.

The transmission function is obtained as ascii-file trm.dat. The energy scale is kinetic energy.

 

Measurement conditions:

XPS measurements were performed at an Axis Ultra DLD (KRATOS, Manchester, UK) with monochromatic Al K radiation (E = 1486.6 eV). The electron emission angle was 0° and the source-to-analyzer angle was 60°. The binding energy scale of the instrument was calibrated following a Kratos analytical procedure, which uses ISO 15472 binding energy data. The setting of the instrument was the hybrid lens mode and the slot mode with an analysis area of approximately 300x700 m². Furthermore, charge neutralization with a flood gun was used. All spectra were recorded in the fixed analyzer transmission (FAT) mode. The samples were measured as powders prepared on a special stainless-steel sample holder.

 

 

 

The data were audited and are BAM reference data.
Files (227.7 kB)
Name Size
jrc620001-c1s.dat
md5:c437b6a720e6b4eea15207b02f30e5d2
28.5 kB Download
jrc620001-xps.vms
md5:b94690976d611db573ff163d7bc07421
73.3 kB Download
jrc620002-c1s.dat
md5:8bb2362d7564013d2979a1cdc605b37b
28.5 kB Download
jrc620002-xps.vms
md5:5bbda6ae83d15dbad62237ff1c890c0e
73.7 kB Download
trm.dat
md5:fde8f3fc983399531f2530528b18ab7d
23.7 kB Download